INTENDED USE:
Disposable Tray Corner Protectors are intended for use with surgical trays to prevent the tearing of the sterilization wrap and to protect the corners of the trays from damage.
INSTRUCTIONS FOR USE:
The corner protectors do not need to be cleaned or disinfected prior to use.
1. Place the corner protectors on the top and bottom corners of the tray before wrapping them with sterilization wrap. Once in place, wrap the tray according to standard wrapping procedures.
2. The corner protectors can be sterilized according to the parameters detailed below. The corner protectors and tray surface under the corner protector will be sterile.
3. The corner protectors are designed for one-time use and should be disposed of after use. There are no special disposal instructions.
PRODUCT SPECIFICATION:
MATERIAL | closed-cell polyurethane ether |
DIMENSION | 6*6*1.25” |
WEIGHT | 12G |
STERILIZATION:
Appropriate sterilant contact will occur on the surface of the corner protectors as well as the tray surface under the corner protectors:
Cycle Type | Temperature | Minimum Time –
Full Cycle |
Pre-Vacuum | 132° C
(270° F) |
4 minutes |
Gravity | 121° C
(250° F) |
30 minutes |